This temperature Calibrator gives high accuracy of 0.02% for measure and source. It has dual display, 5 digits for measure and 6 digits for source & white LED for backlight. Volts, RTD & Thermocouple. It gives output up to 500 to 5k and output for calibrating R, S, K, E, J, T, N, B, L, U type of Thermocouples RTD & Continuity.

Request Callback

Order In Bulk

Generate Quote

100% ORIGINAL Products
100% ORIGINAL Products Details

100% ORIGINAL Products

SECURE Payments
SECURE Payments Details

SECURE Payments

₹ 75,999.00

(GST inclusive price) i
  • Base Value: ₹ 64405.9
  • GST: (18) % ₹ 11593.1
  • Total Value ₹ 75999

₹98,400.00

( - 23% OFF )

Need Help ?

Our Client Care Managers Are On Call 24/7 To Answer Your Question.

Description

Key Features

  • High Accuracy of 0.02% for measure and source.
  • 5 digits display for measure.
  • 6 digits display for source.
  • A dual, Liquid Crystal Display.
  • White LED for backlight

 

What's in the box?

  • Kusam Meco KM CAL 906 Temperature Calibrator
  • Test leads
  • User’s Manual
  • Holster
  • Carrying Case.

Rating & Reviews

4.5
(Based On 10 Reviews)

Rating

0

0

0

0

0
Review Image

Product Specifications

GENERAL

Accuracy : ± 0.02%
Max. Allowable Voltage : 30V
Operation Temperature Range : 0 ~ 50°C
Operation Humidity Range : 80% RH
Storage Temperature Range : -10°C ~ 55°C
Storage Humidity Range: 90% RH
Power supply :
Size : 205 x 95 x 42 mm
Weight : 500g (with holster)
Safety : Conforms to IEC 1010

 

TECHNICAL

Measuring FunctionRangeMeasuring RangeResolutionAccuracyRemark
DC Voltage50mV5.000mV ~ 55.000mV1 V0.02 + 0.02Input Resistance : approximately 100M
500mV-50.00mV ~ 550.00mV10 V0.02 + 0.01
Resistance (OHM)5000.00 ~ 550.000.010.05 + 0.02Test current for 500 : Approximately 1mA; Test current for 5K : Approximately 0.1mA; Open circuit voltage: About 2.5V; Does not include lead resistance:
5K 0.0000K ~ 5.5000K0.10.05 + 0.02
ThermocoupleR0 C ~ 1767 C1 C0~500 C : 1.8 CBy using ITS-90 temperature scale: The accuracy does not include the error of internal temperature compensation caused by a sensor;
S0 C ~ 1767 C500~1767 C : 1.5 C
K-100.0 C ~ 1372.0 C0.1 C-100.0~0.0 C : 1.2 C
0.0~1372.0 C : 0.8 C
E-50.0 C ~ 850.0 C-50.0 C~0.0 C : 0.9 C
0.0 C~85.0 C : 1.5 C
J-60.0 C ~ 1120.0 C-60.0~0.0 C : 1.0 C
0.0~1120.0 C : 0.7 C
T-100.0 C ~ 400.0 C-100.0 ~ 0.0 C : 1.0 C
0.0~400.0 C : 0.7 C
N-200.0 C ~ 1300.0 C-200.0~0.0 C : 1.5 C
0.0~1300.0 C : 0.9 C
B600 C ~ 1820 C1 C600~800 C : 2.2 C
800 ~1000 C : 1.8 C
1000 ~1820 C :1.4 C
L-60.0 C ~ 900.0 C0.1 C-60.0 ~ 0.0 C : 0.7 C
0.0 ~ 900.0 C : 0.5 C
U60.0 C ~ 900.0 C0.1 C-100.0 ~ 0.0 C :0.7 C
0.0 ~ 600.0 C : 0.5 C
Thermo Resistance (RTD) ContinuityPT100 385200.0 C ~ 800.0 C0.1 C-200.0 ~ 0.0 C :0.5 CBy using temperature scale ITS-90. Does not include lead resistance. Assuming all three RTD leads are matched for 3-w input.
0.0 ~ 400.0 C : 0.7 C
400.0~800.0 C:0.8 C
PT 1000 385200.0 C ~ 630.0 C200.0~100.0 C:0.3 C
100.0~300.0 C:0.5 C
300.0~630.0 C:0.7 C
PT 200 385200.0 C ~ 630.0 C-200.0~100.0 C:0.8 C
100.0~300.0 C:0.9 C
300.0~630.0 C:1.0 C
PT 500 385200.0 C ~ 630.0 C-200.0~100.0 C:0.4 C
100.0~300.0 C:0.5 C
300.0~630.0 C:0.7 C
Cu10-100.0 C ~ 260.0 C1.8 C
Cu50 -50.0 C ~ 150.0 C0.7 C
500≤ 500.01 Approximately 1mA test Current