Item(s) Added To Your Bag

Metravi PM-39 Process Meter

Click To Zoom

Metravi PM-39 Process Meter

Metravi PM-39 Process Meter

(Based On 10 Reviews)

5 Digits, Digital Multimeter combined with source for DC voltage, resistance, frequency, thermocouples, thermal resistance, and DC current. supply power to 24V circuits and meanwhile measure current; with built-in250 Ω HART loop resistance

Request Callback

Order In Bulk

Generate Quote

100% ORIGINAL Products
100% ORIGINAL Products Details

100% ORIGINAL Products

SECURE Payments
SECURE Payments Details

SECURE Payments

₹ 51,999.00

(GST inclusive price) i
  • Base Value: ₹ 44066.9
  • GST: (18) % ₹ 7932.1
  • Total Value ₹ 51999

₹69,863.00

( - 26% OFF )

Need Help ?

Our Client Care Managers Are On Call 24/7 To Answer Your Question.

Description

Key Features

  • Measurement Functions: Measurement of AC Voltage, DC Voltage, Ohm, Capacitance, DC Current, AC Current, On-Off, Diodes, Frequency, Thermocouples, Thermal Resistance; Data display and Retention; Measurement of relative values
  • Output Functions: Output of DC voltage, Resistance, Frequency, Thermocouples, Thermal Resistance, and DC Current (constant output, manual stepping and SIMULATE);
  • Loop Inspection: Supply power to 24V circuits and meanwhile measure current; with built - in 250Ω HART loop resistance.

 

What's in the box?

  • Metravi PM-39 Process Meter+
  • One pair of testing wires (including analligator clip)
  • One operating manual
  • Four 1.5V alkaline cells (LR6)
  • One 500mA/250V quick-acting fuse
  • One soft portable bag

Rating & Reviews

4.5
(Based On 10 Reviews)

Rating

0

0

0

0

0
Review Image

Product Specifications

GENERAL

DisplayDigit: 4-digit display (for current measurement and output: 5-digit display)
Display refreshingFast (F): 20 times/second; slow (S): 5 times/second
Temperature and humidity range for work0 ~ 40 ºC, relative humidity ≤85% (without moisture condensation)
Temperature and humidity range for storage-20 ºC ~ 60 ºC, relative humidity below 90%
(without moisture condensation)
Temperature and humidity range for guaranteed precision23 ± 5ºC, relative humidity below 75% (without moisture condensation)
Temperature factor0.1× basic precision / °C (temperature range: <18°C or >28°C)
Application environmentIndoors, outdoors (non-watertight), altitude of 0 ~ 2000m
Indication of outrangeOL
On-Off / open-circuit testBuzzer beeps indicate the resistance reading is lower than the threshold, or an open circuit
Battery typeFour 1.5V (LR6) alkaline batteries
Service life of batteriesWhen alkaline batteries are used,
For measurement of all parameters: about 100 hours
For DC current output (SIMULATE): about 50 hours
For DC current output (SOURCE) 20mA (load of 1000Ω): about 2.5
hours 
Low battery indicationIt is indicated with a battery mark.
Automatic shutdownThe meter is automatically shut down after about 5 minutes of no operation. The time can be adjusted.
Warm-up time10 minutes
Close meter calibrationNo need for internal adjustment
Battery coverFor battery replacement, without influencing meter calibration
Size206 (L)×97 (W)×60 (D)mm
WeightAbout 500g

 

TECHNICAL

FunctionRangeMeasuring scopeResolutionPrecision
DC voltage DCV50m-55.00mV ~ 55.00mV0.01mV0.1%+4
500mV-550.0mV ~ 550.0mV0.1mV0.1%+4
5V-5.500V ~ 5.500V0.001V0.1%+4
50V-55.00V ~ 55.00V0.01V0.1%+4
500V-550.0V ~ 550.0V0.1V0.1%+4
1000V-1000V ~ 1000V1V0.1%+4
AC voltage ACV5V0 ~ 5.500V0.001V0.5%+4(<400Hz)
5%+4(>400Hz)
50V0 ~ 55.00V0.01V 0.5%+4
500V0 ~ 550.0V0.1V0.5%+4
1000V0V ~ 750V1V0.5%+4
OHM ( Ω )500Ω0 ~ 550.0Ω0.1Ω0.1%+4
5KΩ0 ~ 5.500KΩ0.001KΩ0.1%+4
50KΩ0 ~ 55.00KΩ0.01KΩ0.1%+4
500KΩ0 ~ 550.0KΩ0.1KΩ0.5%+4
5MΩ0 ~ 5.500 MΩ0.001MΩ1%+4
50MΩ0 ~ 55.00 MΩ0.01MΩ1%+4
DC current DCI50mA-55.000mA ~ 55.000mA0.001mA0.1%+5
500mA-500.00mA ~ 500.00mA0.01mA0.1%+5
AC current ACI50mA0000mA ~ 55.000mA0.001mA0.5%+10
500mA0.00mA ~ 500.00mA0.01mA0.5%+10
Frequency FREQ10Hz0 ~ 9.9999Hz0.0001Hz0.02%+4
 100Hz0 ~ 99.999Hz0.001Hz0.02%+4
 1000Hz0 ~ 999.99Hz0.01Hz0.02%+4
 10kHz0 ~ 9.9999kHz0.0001kHz0.02%+4
 100kHz0 ~ 99.999kHz0.001kHz0.02%+4
 DUTY10% ~ 90%0.1%1%
Diode2V 0.001V1%+10
On-off test500Ω 0.1Ω≤50ΩBB

 

Detailed precision indexes for measurement (Contd.)

Thermocouple TCR-40°C ~ 1760°C1°C0.5%+3°C (≤100) °C
0.5%+2°C(>100) °C
S-200°C ~ 1760°C
B400°C ~ 1800°C
K-200°C ~ 1350°C0.5%+2°C (≤-100) °C
0.5%+1°C (>-100) °C
E-200°C ~ 700°C
J-200°C ~ 950°C
T-200°C ~ 400°C
N-200°C ~ 1300°C 
Thermal resistance RTDCu50-50°C ~ 150°C1°C0.5%+3°C
 
 Pt100-200°C ~ 850°C
 
 
Capacitance
CAP
10nF0 ~ 11.00nF0.01nF5%+50
100nF0 ~ 110.0nF0.1nF5%+5
1000nF0 ~ 1100nF1nF5%+5
10µF0 ~ 11.00µF0.01uF5%+5
100µF0 ~ 110.0µF0.1uF5%+5
1000µF0 ~ 1100µF1uF5%+5
10mF0 ~ 11.00mF0.01mF5%+50
100mF0 ~ 110.0mF0.1mF5%+50

 

Detailed precision indexes for output

FunctionRangeOutput setting scopeResolutionPrecisionRemark
DC voltage DCV100mV-10.00 ~ 110.00mV10μV0.2%+4Maximum output
current 0.5mA
1000mV-100.0 ~ 1100.0mV100μV0.2%+4Maximum output
current 2mA
10V-1.000 ~ 11.000V1mV0.2%+4Maximum output
current 5mA
DC current DCI30mA0.000 ~ 33.000mA0.001mA0.2%+420mA, maximum load1KΩ 30mA, maximum load600Ω
Simulated transmitter SIMULATE-30mA0.000 ~ -33.000mA 0.001mA
Loop power LOOP24V  ±10%Maximum output current 35mA
OHM400Ω0.0Ω ~ 400.0Ω0.1Ω0.2%+4Excitation current: ±0.5 ~ 3mA When excitation current is ±0.1 ~ 0.5mA, an additional error of 0.1Ω should be taken into account. The precision doesn’t include lead resistance.
Thermocouple TCR0°C ~ 1767°C1°C0.2%+3°C (≤100)
0.2%+2°C (>-100)
With the thermometric
scale of ITS-90;
The precision doesn’t
include errors in cold-end
compensation
S0°C ~ 1767°C
B600°C ~ 1820°C
K-200.0°C ~ 1372.0°C0.1°C0.2%+2°C (≤-100)
0.2%+1°C (>-100)
E-200.0°C ~ 1000.0°C
J-200.0°C ~ 1200.0°C
T-250.0°C ~ 400.0°C
N-200.0°C ~ 1300.0°C
Thermal resistance RTDPT100-0200.0 ~ 0850.00.1°C0.2%+0.6°CWith the thermometric scale of Pt100-385; With excitation current of ±1mA The precision doesn’t include lead resistance.
Cu50-050.0 ~ 150.0
Frequency FREQ100Hz1.0Hz ~ 110.0Hz0.1Hz0.2%+2Rectangular wave, duty
cycle of 50%
1 ~ 11Vp-p
1KHz0.100KHz ~ 1.100KHz 1Hz 0.2%+2
10KHz1.0KHz ~ 11.0KHz 0.1KHz0.2%+2

 

Input characteristics

Function positionInput impedance (nominal value)
V10MΩ, <100pF
mV >2.5GΩ
  
 Common-mode rejection ratioSeries-mode rejection ratio
DCV, DCmV80dB (dc to 50Hz / 60Hz/1KΩ)40dB (50Hz / 60Hz)
ACV, ACmV60dB (dc to 50Hz / 60Hz/1KΩ) 
 Open-circuit voltageFull-scale voltage
Ohm2.5V2.2V
Diode< 3.5V2.2V
On-off< 1V500mV
 Typical short-circuit current
Ohm500Ω5KΩ50KΩ500KΩ5MΩ50MΩ
 0.8mA0.2mA20μA2μA0.2μA< 0.1μA
Diode0.2mA (typical value)

Rating & Reviews

4.5
(Based On 10 Reviews)

Rating

0

0

0

0

0
Review Image