Item(s) Added To Your Bag

Fluke 961C Temperature Humidity Logger with USB Connection

Click To Zoom

Fluke 961C Temperature Humidity Logger with USB Connection

Fluke 961C Temperature Humidity Logger with USB Connection

(Based On 10 Reviews)

No additional data cable is required, by simply using the USB interface of the Fluke 961, you can easily read temperature and humidity report in PDF generated automatically, and you can also quickly export the temperature and humidity data in the Logger in the form of tables, charts, and text. Up to 32,000 sets of data for long-term logging without exporting frequently. The Fluke 961 Series Temperature Humidity Logger with USB Connection is a convenient tool for monitoring temperature and humidity in food, pharmaceutical, cold-chain and cold storage industries.

Request Callback

Order In Bulk

Generate Quote

100% ORIGINAL Products
100% ORIGINAL Products Details

100% ORIGINAL Products

SECURE Payments
SECURE Payments Details

SECURE Payments

₹ 6,189.00

(GST inclusive price) i
  • Base Value: ₹ 5244.9
  • GST: (18) % ₹ 944.1
  • Total Value ₹ 6189

₹8,400.00

( - 26% OFF )

Need Help ?

Our Client Care Managers Are On Call 24/7 To Answer Your Question.

Description

Key Features 

  • USB interface for direct reading PDF report, no software required.
  • 32,000 sets of data for long time logging.
  • Minimum logging interval of 10 seconds, instantaneous logging of rapid environment changes.
  • FDA compliant, data traceable.
  • Multiple formats for data export and reporting.
  • 1 meter drop test, IP54 ingress protection.
  • Compact and light, easy to record and read data at any time.

 

What's in the box?

  • Fluke 961C Temperature Humidity Logger
  • Mounting bracket and screws
  • Velcro hook and loop fastener

Rating & Reviews

4.5
(Based On 10 Reviews)

Rating

0

0

0

0

0
Review Image

Product Specifications

Sensor Type External
Temperature Measurement
Temperature Measurement Range -40.0 °C to 85.0 °C
Accuracy± 0.4 °C
Resolution 0.1 °C
Humidity Measurement
Range0% to 100.0% RH
Accuracy±3% RH (20 to 80% RH @ 25 °C); ±5% RH (<20% RH, >80% RH @ 25 °C)
Resolution0.1% RH
Date Storage32 000
Logging Interval10 s to 24 hr
Start Mode Press buon or use software
Alarm ThresholdUp to 3 upper temperature limits and 2 lower temperature limits
 Alarm Type Single / Cumulative
Report Format PDF (direct read via a PC), Excel and TXT (via PC software)
General Specifications
TemperatureOperating-30 °C to 70 °C (the Logger)
Storage-30 °C to 70 °C (the Logger)
AltitudeOperating2 000 m
Storage12 000 m
Relative HumidityOperating0% RH to 100% RH, non-condensing
Storage<75% RH
Power SupplyBaery type CR2450 x 1
Baery Life 1 year (logging @ 15 min interval, 25 °C)
Safety IEC 610101-1, Pollution Degree 2
Dimensions (L x W x H) 107 mm x 47 mm x 17 mm (No exteal probe)
Weight78 g (with baery)
IP Rating IEC 60529: IP65
Electromagnetic Compatibility (EMC) IEC 61326-1: Basic Electromagnetic Environment; CISPR 11: Group

Rating & Reviews

4.5
(Based On 10 Reviews)

Rating

0

0

0

0

0
Review Image