Shop By Categories
© 2023.Insuzo. All Rights Reserved.
Electrical and Electronic
Enviornmental
General Laboratory | Analytical
Imaging Instruments
Mechanical and Quality Control
Time, Speed and Dimension
Calibrators
Shop By Industry
Item(s) Added To Your Bag
MECO PHA 5850A Power & Harmonics Analyzer
MECO PHA 5850A Power & Harmonics Analyzer
(Based On 10 Reviews)
MECO Power and Harmonics Analyser Model PHA-5850 can Analyze, Measure, Monitor &Data Log values of Power Quality & Consumption (Energy).
Capable of analyzing IT standby power consumption to the maximum demand of factory. It comes with a user friendly application software that increases
the utility & performance of this instrument. The analyzer is ideal for an any Engineer / Inspector for carrying out Periodic Visits, Maintainance of Plant, Vigilance
checks, Surveys and Energy Audits for checking at Industrial and Consumers end.
Request Callback
Order In Bulk
Generate Quote
₹ 2,62,881.00
(GST inclusive price) i₹2,62,881.00
( - 0% OFF )
Description
What's in the box?
|
Rating & Reviews
Rating
General Specification | |
Power | Eight x 1.5V AA Batteries |
External DC Input | Power supply adapter 12 Volts. DC |
Display | Dot Matrix LCD (240x128) with backlight |
LCD Update Rate | 1 time / second |
Power Consumption | 140mA (approx.) |
No. Of Samples | 1024 samples / period |
Data Logging Files | 85 |
Max. File Capacity | 17474 records (3P4W, 3P3W) 26210 records (1P3W) 52420 records (1P2W) 4096 records (50 Harmonics / record) 2 to 3000 seconds for data logging |
Sampling Time | OL |
Operating Condition | -10°C to 50°C < 85% RH |
Storage Condition | -20°C to 60°C < 75% RH |
Dimensions | 257 x 155 x 57 mm |
Weight | 1160 g (Batteries included) |
Model : PHA-5850A (100A) (Overload Protection AC 200A) | ||
Range | Resolution | Accuracy of Readings |
0.04 - 1A | 0.1mA / 1mA | ± 0.5% ± 0.05A |
0.4 - 10A | 0.001A / 0.01A | ± 0.5% ± 0.05A |
4 - 100A | 0.01A / 0.1A | ± 1.0% ± 0.5A |
Model : PHA-5850B (1000A) (Overload Protection AC 2000A) | ||
Range | Resolution | Accuracy of Readings |
10.00A | 0.001A / 0.01A | |
5A - 100.0A | 0.01A / 0.1A | ± 0.5% ± 0.5A |
50A - 1000.0A | 0.1A / 1A | ± 0.5% ± 5A |
Model : PHA-5850C (3000A) (Overload Protection AC 3000A) | ||
Range | Resolution | Accuracy of Readings |
10.0 - 300.0A | 0.01A / 0.1A | ± 1% of range |
300.0 - 3000A | 0.1A / 1A |
Model : PHA-5850D (1200A) (Overload Protection AC 1200A) | ||
Range | Resolution | Accuracy of Readings |
6.0 - 120.0A | 0.01A / 0.1A | ± 1% of range |
120.0 - 1200A | 0.1A / 1A |
Harmonic of AC Voltage in Percentage | ||
Range | Resolution | Accuracy of Readings |
1 - 10th | 0.1% | ± 2% |
11 - 20th | ± 4% of reading ± 2.0% | |
50 - 99th | ± 6% of reading ± 2.0% |
Harmonic of AC Current in Percentage Model : PHA-5850A (100A) | ||
Range | Resolution | Accuracy of Readings |
1 - 10th | 0.1% | ± 0.2% of reading ± 1% |
11 - 20th | ± 2% of reading ± 1% | |
21 - 50th (A range) | ± 5% of reading ± 1% | |
21 - 50th (mA range) | ± 10% of reading ± 1% | |
51 - 99th | ± 35% of reading ± 1% |
Model : PHA-5850B (1000A) | ||
Range | Resolution | Accuracy of Readings |
1 - 20th | 0.1% | ±2% |
21 - 49th | ± 4% of reading ± 2.0% | |
50 - 99th | ± 6% of reading ± 2.0% |
Model : PHA-5850C (3000A) & PHA-5850D (1200A) | ||
Range | Resolution | Accuracy of Readings |
1 - 20th | 0.1% | ±2 % |
21 - 49th | ± 6% | |
50 - 99th | ± 10% |
Model : PHA-5850A (100A) | ||
Range | Resolution | Accuracy of Readings |
5.0 - 999.9W | 0.1W | ± 1% ± 0.8W |
1.000 - 9.999KW | 0.001KW | ± 1% ± 8W |
10.00 - 99.99KW | 0.01KW | ± 1% ± 80W |
100.0 - 999.9KW | 0.1KW | ± 1% ± 0.8KW |
1000 - 9999KW | 1KW | ± 1% ± 8KW |
Model : PHA-5850B (100A) | ||
Range | Resolution | Accuracy of Readings |
5.0 - 999.9W | 0.1W | ± 1% ± 0.8W |
1.000 - 9.999KW | 0.001KW | ± 1% ± 8W |
10.00 - 99.99KW | 0.01KW | ± 1% ± 80W |
100.0 - 999.9KW | 0.1KW | ± 1% ± 0.8KW |
1000 - 9999KW | 1KW | ± 1% ± 8KW |
0.000 - 9.999MW | 0.001MW | ± 1% ± 80KW |
Model : PHA-5850C (3000A) Model : PHA-5850D (1200A) | |||
Range | Resolution | Accuracy of Readings | |
10.0 - 999.9W | 0.1W | >20V & >30A | <20V or <30A |
1.000 - 9.999KW | 0.001KW | ± 1% of range | ± 2% of range |
10.00 - 99.99KW | 0.01KW | ||
100.0 - 999.9KW | 0.1KW | ||
1000 - 9999KW | 1KW |
AC Voltage (50Hz or 60H | ||
Range | Resolution | Accuracy of Readings |
20.0V - 500.0V (Phase to Neutral) | 0.1V | ± 0.5% ± 5dgts |
20.0V - 600.0V (Phase to Phase) |
Power Factor (PF) Model : PHA-5850A (100A) & PHA-5850B (1000A) | ||
Range | Resolution | Accuracy of Readings |
0.00 - 1.00 | 0.1V | ± 0.04 |
Model : PHA-5850C (3000A) & PHA-5850D (1200A) | |||
Range | Resolution | Accuracy of Readings | |
0.000 - 1.000 | 0.001 | >20V & >30A | <20V or <30A |
± 0.04 | ± 0.1 |
Rating & Reviews
Rating
Similar Products